DECTRIS EIGER2® S photon-counting X-ray detectors are the perfect choice if your research requires large-area, high-resolution beamline detectors, but not the ultimate speed. Like all EIGER2 detectors, this camera with 75-µm pixels delivers data with the highest spatial resolution. It includes two independent energy thresholds for spectral sensitivity, a single-pixel point-spread function for ideal counting statistics, and fast image readout with a frame rate up to 1 kHz. Equipped with a silicon sensor that is 450 µm thick, EIGER2 S detectors cover X-ray energies between 6 and 40 keV.
EIGER2 S detectors come in a variety of sizes. Their robustness, combined with excellent support, make EIGER2 S detectors the most reliable and future-proof tool for cost-conscious beamlines.
-
More & better data: these detectors have a large active area, two adjustable energy-discriminating thresholds, and a high count rate (over 107 ph/s/pixel). There is no readout noise or dark current.
- Affordable: EIGER2 S detectors offer high resolution for an unbeatable price per pixel.
- Reliable: these detectors feature extremely stable performance and radiation hard design. Delivery terms are guaranteed, thanks to streamlined industrial production and top-notch support over the product’s entire lifetime.
Detectors
EIGER2 S 16M
This is the largest DECTRIS detector with 75 μm pixel size, ideal for macromolecular crystallography with high count rate at a modest price tag.
EIGER2 S 9M
This detector with more than nine million pixels, is an excellent and affordable allrounder for macromolecular crystallography and other synchrotron applications that demand highest pixel resolution.
EIGER2 S 4M
This detector is an excellent tool for diffraction, SAXS, or ptychography in cases where ultimate speed is not required. In-vacuum compatability is optionally available.
EIGER2 S 1M
With a frame rate of 1,000 Hz and an extremely high count-rate capability, this is a universal and affordable tool for scattering or diffraction experiments.
Detector Specifications*
EIGER2 S 16M | EIGER2 S 9M | EIGER2 S 4M | EIGER2 S 1M | |
---|---|---|---|---|
Active area (W x H) [mm² ] |
311.1 x 327.2 |
233.1 x 244.7 |
155.1 x 162.2 |
77.1 x 79.7 |
Pixel array (W x H) |
4,148 x 4,362 |
3,108 x 3,262 |
2,068 x 2,162 |
1,028 x 1,062 |
Pixel size (W x H) [µm² ] |
75 x 75 |
75 x 75 |
75 x 75 |
75 x 75 |
Inter-module gap (Hor./Vert.) [pixels ] |
12 / 38 |
12 / 38 |
12 / 38 |
- / 38 |
Count rate (max.) [ph/s/pixel ] |
10⁷ |
10⁷ |
10⁷ |
10⁷ |
Energy range [keV ] |
6 - 40 |
6 - 40 |
6 - 40 |
6 - 40 |
Number of energy thresholds |
2 |
2 |
2 |
2 |
Threshold range [keV ] |
3.5 - 30 |
3.5 - 30 |
3.5 - 30 |
3.5 - 30 |
Frame rate (max.)1 [Hz ] |
25 |
40 |
100 |
1,000 |
Region of Interest (ROI) readout |
Yes |
Yes |
Not applicable |
Not applicable |
Readout time |
Continuous readout |
Continuous readout |
Continuous readout |
Continuous readout |
External gating |
Not applicable |
Not applicable |
Not applicable |
Not applicable |
Image bit depth [bits ] |
32 |
32 |
32 |
32 |
Sensor material |
Silicon (Si) |
Silicon (Si) |
Silicon (Si) |
Silicon (Si) |
Sensor thickness [µm ] |
450 |
450 |
450 |
450 |
Point-spread function (FWHM) [pixels ] |
1 |
1 |
1 |
1 |
Vacuum compatibility |
Not applicable |
Optional |
Optional |
Optional |
Dimensions (W x H x D) [mm³ ] |
400 x 430 x 500 |
340 x 370 x 500 |
235 x 237 x 372 |
114 x 133 x 242 |
Weight [kg ] |
55 |
41 |
15 |
4.7 |
Cooling method |
Water (Module) / Air (Electronics) |
Water (Module) / Air (Electronics) |
Water (Module) / Air (Electronics) |
Water (Module) / Air (Electronics) |
|
1 One frame (16 bits) per exposure, i.e. single-threshold or difference image.
2 Effective dead time between exposures < 100 ns (max. loss of 1 count/pixel)
*All specifications are subject to change without notice.
Key Contact
Request a quote
To get more information for a tender or learn more about purchasing a DECTRIS EIGER2® X-ray detector for synchrotron beamlines , reach out to us.
Want to know more?
We are happy to answer any questions about us, our detectors, or the process of integrating our detectors into your experimental setup.
Need support?
Our Support team is here to answer your questions and help you take full advantage of your detector.